Make : EURO EA
Model : Euro EA3000
Specification : Estimation of CHN/CHNS/O in percentage
  level to high concentration level.

         EuroEA3000 is a series of compact, fully automatic Elemental Analysers for the most exacting determination of CHNS and O. It is based on TurboFlash Combustion Technology, an innovative principle for pressurised oxygen injection and represents a quantum leap in state of the art of dynamic flash combustion as well as High Temperature (HT) pyrolytic Oxygen and Hydrogen analysis for elemental and isotopic determination.


 
 
  Working Principles   
 

         EuroEA3000 Series offers independently programmed oxygen volume and oxygen pressure: the "Pressurized Oxygen Injection System" (TurboFlash Combustion Technology) provides full flexibility and control of the combustion process. Catalysts and oxygen are conserved, difficult matrices are broken with ease and analytical precision dramatically improved for both elemental and isotopic determinations.

• EuroEA3000 is a rapid CHNS-O Analyser delivering the highest levels of speed and performance yet achieved. Built-in leak detection, wake up routine, flow rate monitoring, quick and safe exhausted reactor change, unlimited methods storage of pre-set parameters, are a few of the new outstanding features.
• Instrument operation is truly intuitive for the first time through the use of Callidus SW allowing full automation and manual override at the instrument keypad. Callidus sets all analytical parameter, integrates and reprocesses data according to GLP, prompts routine maintenance and catalyst identification with specific part numbers shown on screen.
• The same basic instrument can be configured for all current and emerging new applications with the largest number of options. EuroEA3000 delivers maximum flexibility through modularity and upgradability.
• EuroEA3000 with new long life catalysts designed for pressurized oxygen injection is cost effective to operate with unprecedented reliability and minimum downtime. Gas saving routine with zero gas flow in stand-by and intelligent catalyst protection are included.
            Key components of CHNS are auto sampler, combustion reactors ,chromatographic column, and T.C.D. detector.


 
  Applications   
     Soil Chemical Analysis

 
  User Instructions   
         1. Sample must be pure and all contaminants must be removed.
        2. Sample weight should be in between 4-5 mg.
        3. Liquid samples should have constant weight.

 
 
  Contact Us   


         Director, ICAR-IIFSR

         Contact : 0121-2888711
         Email Id: director.iifsr@icar.gov.in

 

 


 
 CHNS (O) Analyzer Charges includes GST  
 
  Industry University National Lab/R&D's  
CHNS Estimation 2360 /- 443/- 1062/- Per Sample

*These are tentative rates, the actual rates will be decided by the Committee soon.

kodowanie html5 css, projektowanie stron internetowych , PSD2HTML, PSD HTML, proste strony www, cięcie PSD, tworzenie stron,